Brand:Ainuo
Description:4 in 1 Safety Tester AC/DC Hipot, Insulation Resistance, Ground Bond, Wait; Safety accuracy: 2%; AC vector compensation, DC fast discharge, arc detection; Support barcode scanner, U Disk
Origin:China Shandong
Model:AN9637H(F)/AN9638H(F)
FOB:perUnit/Units500--2000USD
Payment Terms:Other
Supply Ablility:200Unit/Units
Supply Time:Month
Packaging Details:Wooden case
Product detail:
AC Hipot test |
||
Rated output capacity |
AN9638H(F) |
500VA(5000V/100mA),short circuit current >200mA |
AN9637H(F) |
200VA(5000V/40mA),short circuit current >200mA |
|
Output voltage setting |
Range,Accuracy |
100~5000V,±(2%×setting+5counts) |
Output frequency setting |
Range,Accuracy |
50Hz or 60Hz,±0.1%×setting |
Alarm current limits setting |
Upper-limit range,Accuracy |
0.10~100.00mA(or 40.00mA),±(2%×setting+5counts) |
Lower-limit range,Accuracy |
0.000~9.999mA,±(2%×setting+5counts) |
|
Time setting |
Range |
0.5~999.9s(0 is infinite) |
Ramp up/down time range |
0.1~999.9s |
|
Accuracy |
±(0.2%×setting+1count) |
|
Measurement |
Voltage range,Accuracy |
0.01~5.00kV,±(1.5%×reading+1count) |
Frequency range,Accuracy |
50Hz or 60Hz,±0.1%×reading |
|
Current range,Accuracy |
0.10~100.00mA(or 40.00mA),±(2%×reading+2counts) |
|
Time range,Accuracy |
0.1~999.9s,±(0.2%×reading+1count) |
|
Output voltage waveform, distortion, regulation |
Sine wave,<2%(pure resistance load),<2%×setting+5V)(from empty load to full load) |
|
Start voltage setting |
(0%~50%)×output voltage setting |
|
Current compensation setting |
0.000~100.00mA(or 40.00mA),auto、manual |
|
Arc detection |
1~9 level,9 level is optimally sensitive,0 is OFF |
|
DC Hipot test |
||
Output voltage setting |
Range,Accuracy |
100~6000VDC,±(2%×setting+5counts) |
Alarm current limits setting |
Upper-limit,Accuracy |
0~10000μA,±(2%×setting+5counts) |
lower-limit,Accuracy |
0.0~999.9μA,±(2%×setting+5counts) |
|
Time setting |
Range |
0.5~999.9s(0 is infinite) |
Ramp up time range |
0.1~999.9s |
|
Ramp down time range |
0.3~999.9s(0 is OFF) |
|
Accuracy |
±(0.2%×setting+1 count) |
|
Measurement |
Voltage range,Accuracy |
0.01~6.00VDC,±(1.5%×reading+1count) |
Frequency range,Accuracy |
0.0~10000uA,±(2%×reading+5count) |
|
Current range,Accuracy |
0.1~999.9s,±(0.2%×reading+1count) |
|
Rated output capacity |
60VA(6000VDC/10mA) |
|
Output voltage ripple, regulation |
<5%(pure resistance load,6000V/10mA),<(2%×setting+5V)(from empty load to full load) |
|
Compensation current Setting |
0 ~200.0μA,auto, manual |
|
Start voltage setting |
(0%~50%)× setting output voltage |
|
Charging current lower limit |
(0~3.500)μA,auto, manual |
|
Discharge time |
≤200ms |
|
Arc detection |
1~9 level,9 level is optimally sensitive,0 is OFF |
|
Insulation Resistance test |
||
Output voltage setting |
Range,Accuracy |
100~2500Vdc,±(2%×setting+5counts) |
Alarm resistance setting |
Range |
1~9999MΩ |
Accuracy |
100-499V,1~1000MΩ,±(8%×setting+2counts) 1000~2000MΩ,±(12%×setting+2counts) 500-2500V,1~199MΩ,±(2%×setting+2counts), 200~999MΩ, ± (5%×setting+2counts) 1000~9999MΩ, ± (15%×setting+2counts) |
|
Time setting |
Ramp up time range |
0.1~999.9s |
Adjustment delay time |
0,0.5~999.9s(0 is infinite) |
|
Test time range |
0,0.5~999.9s(0 is infinite) |
|
Ramp down time range |
0.1~999.9s |
|
Accuracy |
±(0.1%×setting+0.1s) |
|
Measurement |
Voltage range, accuracy |
100~2500Vdc,±(2%×reading+5counts) |
Resistance range, accuracy |
1~9999MΩ; 100-499V,1~1000MΩ,±(8%×reading+2counts) 1000~2000MΩ,±(12%×reading+2counts) 500-2500V,1~199MΩ,±(2%×reading+2counts), 200~999MΩ, ± (5%×reading+2counts) 1000~9999MΩ, ± (15%×reading+2counts) |
|
Time range,accuracy |
0.1~999.9s,±(0.2%×reading+1counts) |
|
Charging current lower limit |
0~3.500μA,auto, manual |
|
Discharge time |
≤200ms |
|
Ground Bond test |
||
Rated output |
AN9638H(F) |
Max output current is 64A,max testing resistance 600mΩ,open-circuit voltage <12V |
AN9637H(F) |
Max output current is 32A,max testing resistance 600mΩ,open-circuit voltage <12V |
|
Output current setting |
Range,Accuracy |
2.0~32.0A(or 64A),±(2%×setting+5counts) |
Output voltage setting |
Range,Accuracy |
3.0~10.0V ,±(2%×setting+5counts) |
Output frequency setting |
Range,Accuracy |
Sine wave,50Hz or 60Hz,±0.1%×setting |
Alarm limits setting |
Resistance upper/lower limits |
2.0~10.6A:0.1~600mΩ; 10.7~32.0A(or 64A):0.1~R mΩ,among R=(6400 / set current value)mΩ,when current range changs,Tester for automatic calculation range scope biggest |
Tolerance |
±(2%×setting+5counts) |
|
Compensation resistance setting |
Range |
0~100 mΩ,automatic measurement,ON/OFF |
Test time setting |
Range,Accuracy |
0.5~999.9s(0 stand for time infinite length);±(0.2%×setting+1count) |
Measurement |
Current range,accuracy |
2.0~32.0A(or 64A),±(2%×reading+2counts) |
Voltage range,accuracy |
3.0~10.0V ,±(2%×reading+5counts) |
|
Resistance range, accuracy |
10.0~99.9~100~600mΩ;±(2%×reading+5counts) (64A current calculated method is same) |
|
Time range,accuracy |
0.5~999.9s(0 stand for time infinite);±(0.2%×setting+1count) |
|
Extended function |
||
Communication function |
Address setting |
0~999 |
Baud rate setting |
300,1200,2400,4800,9600,192000 |
|
Verification mode setting |
None、odd、even |
|
Bar code scanning |
Bar code scanning function |
ON/OFF |
Bar code length setting |
1~30 |
|
Model identification number |
Start bit select 1~30,max length 16 Bit |
|
Function allocation explain |
Hardware interface is optional |
|
print function |
printer type setting |
stylus printer miniprinter |