Safety Tester AN9637H(F)

Click for original picture
Rated output capacity: 500VA(5000V/100mA)
Fob Price: Discuss personally 
Supply Ability: 200
Payment Terms: delivery withindays since payment
Location: China Shandong
  Ada    Contact Now 询价
Not exactly what you want? Post a quick Buying Request !
 
 
Product Detail

Brand:Ainuo

Description:4 in 1 Safety Tester AC/DC Hipot, Insulation Resistance, Ground Bond, Wait; Safety accuracy: 2%; AC vector compensation, DC fast discharge, arc detection; Support barcode scanner, U Disk

Origin:China Shandong

Model:AN9637H(F)/AN9638H(F)

FOB:perUnit/Units500--2000USD

Payment Terms:Other

Supply Ablility:200Unit/Units

Supply Time:Month

Packaging Details:Wooden case

Product detail: 

AC Hipot test

Rated output capacity

AN9638H(F)

500VA(5000V/100mA),short circuit current >200mA

AN9637H(F)

200VA(5000V/40mA),short circuit current >200mA

Output voltage setting

Range,Accuracy

100~5000V,±(2%×setting+5counts)

Output frequency setting

Range,Accuracy

50Hz or 60Hz,±0.1%×setting

Alarm current limits setting

Upper-limit range,Accuracy

0.10~100.00mA(or 40.00mA),±(2%×setting+5counts)

Lower-limit range,Accuracy

0.000~9.999mA,±(2%×setting+5counts)

Time setting

Range

0.5~999.9s(0 is infinite)

Ramp up/down time range

0.1~999.9s

Accuracy

±(0.2%×setting+1count)

Measurement

Voltage range,Accuracy

0.01~5.00kV,±(1.5%×reading+1count)

Frequency range,Accuracy

50Hz or 60Hz,±0.1%×reading

Current range,Accuracy

0.10~100.00mA(or 40.00mA),±(2%×reading+2counts)

Time range,Accuracy

0.1~999.9s,±(0.2%×reading+1count)

Output voltage waveform, distortion, regulation

Sine wave,<2%(pure resistance load),<2%×setting+5V)(from empty load to full load)

Start voltage setting

0%~50%)×output voltage setting

Current compensation setting

0.000~100.00mA(or 40.00mA),auto、manual

Arc detection

1~9 level,9 level is optimally sensitive,0 is OFF

DC Hipot test

Output voltage setting

Range,Accuracy

100~6000VDC,±(2%×setting+5counts)

Alarm current limits setting

Upper-limit,Accuracy

0~10000μA,±(2%×setting+5counts)

lower-limit,Accuracy

0.0~999.9μA,±(2%×setting+5counts)

Time setting

Range

0.5~999.9s(0 is infinite)

Ramp up time range

0.1~999.9s

Ramp down time range

0.3~999.9s(0 is OFF)

Accuracy

±(0.2%×setting+1 count)

Measurement

Voltage range,Accuracy

0.01~6.00VDC,±(1.5%×reading+1count)

Frequency range,Accuracy

0.0~10000uA,±(2%×reading+5count)

Current range,Accuracy

0.1~999.9s,±(0.2%×reading+1count)

Rated output capacity

60VA(6000VDC/10mA)

Output voltage ripple, regulation

<5%(pure resistance load,6000V/10mA),<(2%×setting+5V)(from empty load to full load)

Compensation current Setting

0 ~200.0μA,auto, manual

Start voltage setting

0%~50%)× setting output voltage

Charging current lower limit

0~3.500)μA,auto, manual

Discharge time

≤200ms

Arc detection

1~9 level,9 level is optimally sensitive,0 is OFF

Insulation Resistance test

Output voltage setting

Range,Accuracy

100~2500Vdc,±(2%×setting+5counts)

Alarm resistance setting

Range

1~9999MΩ

Accuracy

100-499V,1~1000MΩ,±(8%×setting+2counts)

1000~2000MΩ,±(12%×setting+2counts)

500-2500V,1~199MΩ,±(2%×setting+2counts),

200~999MΩ, ± (5%×setting+2counts)

1000~9999MΩ, ± (15%×setting+2counts)

Time setting

Ramp up time range

0.1~999.9s

Adjustment delay time

0,0.5~999.9s(0 is infinite)

Test time range

0,0.5~999.9s(0 is infinite)

Ramp down time range

0.1~999.9s

Accuracy

±(0.1%×setting+0.1s)

Measurement

Voltage range, accuracy

100~2500Vdc,±(2%×reading+5counts)

Resistance range, accuracy

1~9999MΩ;

100-499V,1~1000MΩ,±(8%×reading+2counts)

1000~2000MΩ,±(12%×reading+2counts)

500-2500V,1~199MΩ,±(2%×reading+2counts),

200~999MΩ, ± (5%×reading+2counts)

1000~9999MΩ, ± (15%×reading+2counts)

Time range,accuracy

0.1~999.9s,±(0.2%×reading+1counts)

Charging current lower limit

0~3.500μA,auto, manual

Discharge time

≤200ms

Ground Bond test

Rated output

AN9638H(F)

Max output current is 64A,max testing resistance 600mΩ,open-circuit voltage <12V

AN9637H(F)

Max output current is 32A,max testing resistance 600mΩ,open-circuit voltage <12V

Output current setting

Range,Accuracy

2.0~32.0A(or 64A),±(2%×setting+5counts)

Output voltage setting

Range,Accuracy

3.0~10.0V ,±(2%×setting+5counts)

Output frequency setting

Range,Accuracy

Sine wave,50Hz or 60Hz,±0.1%×setting

Alarm limits setting

Resistance upper/lower limits

2.0~10.6A:0.1~600mΩ;

10.7~32.0A(or 64A):0.1~R mΩ,among R=(6400 / set current value)mΩ,when current range changs,Tester for automatic calculation range scope biggest

Tolerance

±(2%×setting+5counts)

Compensation resistance setting

Range

0~100 mΩ,automatic measurement,ON/OFF

Test time setting

Range,Accuracy

0.5~999.9s(0 stand for time infinite length);±(0.2%×setting+1count)

Measurement

Current range,accuracy

2.0~32.0A(or 64A),±(2%×reading+2counts)

Voltage range,accuracy

3.0~10.0V ,±(2%×reading+5counts)

Resistance range, accuracy

10.0~99.9~100~600mΩ;±(2%×reading+5counts)

64A current calculated method is same)

Time range,accuracy

0.5~999.9s(0 stand for time infinite);±(0.2%×setting+1count)

Extended function

Communication function

Address setting

0~999

Baud rate setting

300,1200,2400,4800,9600,192000

Verification mode setting

None、odd、even

Bar code scanning

Bar code scanning function

ON/OFF

Bar code length setting

1~30

Model identification number

Start bit select 1~30,max length 16 Bit

Function allocation explain

Hardware interface is optional

print function

printer type setting

stylus printer miniprinter

   

 
more..Other products from this supplier

[ ProductsSearch ]  [ [ Print ]  [ Close ]


Copyright © 2004 ChinaAseanTrade.com All rights reserved
中国-东盟自由贸易网版权所有@2004 沪lCP证05013507号

沪公网安备 31010102002066号