Add to Favorite | Set Home | Casual look | Mobile

Ainuo Instrument Co.,Ltd.

Company Profile

 Ainuo Instrument Co.,Ltd.

China/Shandong

category
  • No Category
Site Search
 
link
  • No Links
Current:Home » Product Categories » » Safety tester AN1637H(F)
Similar products
Safety tester AN1637H(F)
Click to enlarge
Product:
Brand: Ainuo 
Price: Negotiable 
MOQ:  
Total supply:
Delivery time: From the date of payment from the buyer within days of delivery
update time: 2024-08-05 01:00  Valid to:Long-term effective
  Inquiry
Profile
 
AC withstand voltage test
Rated output capacity 200VA (5000V/40mA), short circuit current is greater than 200mA;
(Optional: 500VA (5000V/100mA), short-circuit current greater than 200mA;)
Output voltage Range/accuracy 100~5000V, ±(1%× setting +5V)
Output frequency Range/accuracy 50Hz or 60Hz, ±0.1%× setting
Alarm current Upper limit/accuracy 0.10~40.00mA, ±(1%× setting +5 words)
Lower limit /accuracy 0.000~9.999mA, ±(1%× setting +5 words)
Time Range 0.5~999.9s (0: infinite time)
Ramp-up time 0, 0.1~999.9s (0: off)
Ramp-down time 0, 0.1~999.9s (0: off)
Accuracy ± (0.2%× setting +1 word)
Measure meter Voltage range/accuracy 0.01~5.00kV, ± (1%× reading +2 words)
Current range/accuracy 0.10~40.00mA, ± (1%× reading +5 words)
Time range/accuracy 0.1~999.9s, ± (0.2%× reading +1 word)
Output voltage wave, distortion, adjusting rate Sine wave, less than 2% (pure resistive load), ±(1% × setting + 5V) (from zero to full load)
Start voltage (0%~50%)× output voltage setting
Current compensation 0.000~40.00mA, Auto/manual
Arc detection Level 1~9, 9: highest level; 0: off
DC withstand voltage test
Output voltage Range/accuracy 100~6000Vdc, ± (1%× setting +5V)
Alarm current Upper limit/accuracy 0~10000μA, ± (1%× setting +5 words)
Lower limit /accuracy 0.0~999.9μA, ± (1%× setting +5 words)
Time Range 0.5~999.9s(0: infinite time)
Ramp-up time 0,0.4~999.9s(0为关闭)
0, 0.4~999.9s (0: off)
Ramp-down time 0, 1.0~999.9s (0: off)
Accuracy ± (0.2%× setting +1 word)
Measure meter Voltage range/accuracy 0.01~6.00Vdc, ± (1%× reading +1 words)
Current range/accuracy 0.0~10000uA, ± (1%× reading +1 words)
Time range/accuracy 0.1~999.9s, ± (0.2%× reading +1 word)
Rated output capacity 60VA(6000Vdc/10mA)
Output voltage ripple, adjustment ± (1%× setting +2V), zero to full load
Compensation current 0 ~200.0μA, Auto/manual
Start voltage (0%~50%)×output voltage setting
Charging current lower limit (0~3.500) μA, Auto/manual
Discharge time ≤200ms
Arc detection Level 1~9, 9: highest level; 0: off
Insulation resistance test
Output voltage Range/accuracy 100~2500Vdc, ± (1%× setting +5V)
Alarm resistance Upper/lower limit 0.1MΩ~999.99MΩ, 1000MΩ~50000MΩ, the upper limit includes no upper limit setting
Accuracy 100-499V, 0.10MΩ~999.99MΩ, 1000~2000MΩ, ± (5%× setting +2 words)
500-2500V, 0.10~999.99MΩ, ± (2%× setting +2 words),
      1000~9999MΩ, ± (5%× setting +2 words)
      10000~50000MΩ, ± (15%× setting +2 words)
Time Ramp-up time 0,0.1~999.9s (0: off)
Delay decision time 0, 0.5~999.9s (0: infinite)
Range 0, 0.5~999.9s (0: infinite)
Ramp-down time 0, 1.0~999.9s (0: off)
Accuracy ± (0.1%× setting +0.1s)
Measure meter Voltage range/accuracy 100~2500Vdc, ± (2%× reading +5 words)
Resistance range/accuracy
 
Range: 0.100 MΩ~9.999 MΩ, 10.00 MΩ~99.99 MΩ, 100.0 MΩ~999.9 MΩ, 1.000 GΩ~9.999 GΩ, 10.00 GΩ~50.00 GΩ
Resolution: 0.001 MΩ, 0.01 MΩ, 0.1 MΩ, 0.001 GΩ, 0.01 GΩ
Resolution:
100-499V, 0.100 MΩ~9.999 MΩ, 10.00 MΩ~99.99 MΩ, 100.0 MΩ~999.9 MΩ, 1.000 GΩ~2.000 GΩ± (5%× reading +2 words)
500-2500V, 0.100 MΩ~9.999 MΩ, 10.00 MΩ~99.99 MΩ, 100.0 MΩ~999.9 MΩ ± (2%× reading +2 words),
 1.000 GΩ~9.999 GΩ, ± (5%× reading +2 words)
 10.00 GΩ~50.00 GΩ, ± (15%× reading +2 words)
Time range/accuracy 0.1~999.9s, ± (0.2%× reading +1 word)
Charging current lower limit 0~3.500μA, Auto/manual
Grounding resistance test
Rated output Maximum output current: 32A; maximum test resistance: 600mΩ; maximum open circuit voltage: 12V
Output current Range/accuracy 2.0~32.0A, ± (1%× setting +2 words)
Output voltage Range/accuracy 3.0~6.4V , ± (1%× setting +2 words)
Output frequency Range/accuracy Sine wave, 50Hz or 60Hz, ±0.1%× setting
Alarm limit Resistance upper/lower limit 2.0~10.6A: 0.1~600mΩ;
10.7~32.0A: 0.1~R mΩ, R= (6400/current setting) mΩ. As the current scale is switched, the tester will automatically calculate the maximum scale. Resolution: 0.1/1 mΩ
Error ± (1%× setting +2 words)
Compensation resistance Range 0~100 mΩ, automatic measurement , compensation On/Off
Test time Range/accuracy 0.5~999.9s(0代表时间无限长);±(0.2%×设定值+1个字)
0.5~999.9s (0: infinite); ± (0.2%× setting + 1 word)
 
 
 
 
 
Measure meter
Current range/accuracy 2.0~32.0A( or 64A), ± (1%× reading +2 words)
Voltage range/accuracy 3.0~10.0V , ± (1%× reading +2 words)
Resistance range/accuracy 10.0~99.9~100~600mΩ;± (1%× reading +2 words)
 
Time  range/accuracy 0.5~999.9s (0: infinite); ± (0.2%× setting +1 word)
General Specification
Verification & Calibration environment 23℃±5℃, (45~75) %RH , less dust
Operating environment 0 ~40℃, , (5~90) %RH , less dust
Storage environment -10~55℃, (5~95) %RH , less dust
Power supply 220V±10%,50Hz±5%
Power consumption Maximum 1000VA (excluding operating power of objects under test)
Weight (excluding packaging)  Approx. 40kg
Dimension (excluding packaging, mm) 483(W)x 178(H)x 550(D
Inquiry list
Related Products